ASTM C1379-97
Historical Standard: ASTM C1379-97 Standard Test Method for Analysis of Urine for Uranium-235 and Uranium-238 Isotopes by Inductively Coupled Plasma-Mass Spectrometry
SUPERSEDED (see Active link, below)
ASTM C1379
1. Scope
1.1 This test method covers the determination of the concentration of uranium-235 and uranium-238 in urine using Inductively Coupled Plasma-Mass Spectrometry. This test method can be used to support uranium facility bioassay programs.
1.2 The method detection limits for 235 U and 238 U are 6 ng/L. To meet the requirements of ANSI N13.30, the minimum detectable activity (MDA) of each radionuclide measured must be at least 0.1 pCi/L (0.0037 Bq/L). The MDA translates to 47 ng/L for 235 U and 300 ng/L for 238 U. Uranium-234 cannot be determined at the MDA with this test method because of its low mass concentration level equivalent to 0.1 pCi/L.
1.3 The digestion and anion separation of urine may not be necessary when uranium concentration of ? ng/L U are present.
Note 1: Warning--The ICP-MS is a source of intense ultraviolet radiation from the radio frequency induced plasma. Protection from radio frequency radiation and UV radiation is provided by the instrument under normal operation.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.
ASTM Standards
C859 Terminology Relating to Nuclear Materials
C1310 Test Method for Determining Radionuclides in Soils by Inductively Coupled Plasma-Mass Spectrometry Using Flow Injection Preconcentration
C1345 Test Method for Analysis of Total and Isotopic Uranium and Total Thorium in Soils by Inductively Coupled Plasma-Mass Spectrometry
D1193 Specification for Reagent Water
Other Documents
DOEOrder5480.11 Radiological Measurements QualityKeywords
bioassay; inductively coupled plasma-mass spectrometry; total uranium concentration; urine analysis; ICP-MS (inductively-coupled plasma-mass spectrometry); Mass spectrometry--nuclear materials/applications; Total uranium concentration; Uranium concentration; Uranium oxide (U 3 O 8 ); Urinalysis;
ICS Code
ICS Number Code 11.100 (Laboratory medicine)
DOI: 10.1520/C1379-97
ASTM International is a member of CrossRef.