ASTM D7237-06
Historical Standard: ASTM D7237-06 Standard Test Method for Aquatic Free Cyanide with Flow Injection Analysis (FIA) Utilizing Gas Diffusion Separation and Amperometric Detection
SUPERSEDED (see Active link, below)
ASTM D7237
1. Scope
1.1 This test method is used to establish the concentration of aquatic 'free' cyanide in an aqueous wastewater or effluent. The test conditions of this method are used to measure free cyanide (HCN and CN - ) and cyanide bound in the metal-cyanide complexes that are easily dissociated into free cyanide ions at the pH of the aquatic environment ranging from pH 6 to pH 8. The extent of HCN formation is less dependent on temperature than the pH; however, the temperature can be regulated if deemed necessary to further simulate the actual aquatic environment.
1.2 The aquatic free cyanide method is based on the same instrumentation and technology that is described in standard test method D 6888, but employs milder conditions (pH 6-8 buffer versus HCl in the reagent stream), and does not utilize ligand displacement reagents.
1.3 The aquatic free cyanide measured by this procedure should be similar to actual levels of HCN in the original aquatic environment. This in turn may give a reliable index of toxicity to aquatic organisms.
1.4 This procedure is applicable over a range of approximately 2 to 500 g/L (parts per billion) aquatic free cyanide. Sample dilution may increase cyanide recoveries depending on the cyanide speciation; therefore, it is not recommended to dilute samples. Higher concentrations can be analyzed by increasing the range of calibration standards or with a lower injection volume. In accordance with Guide E 1763 and Practice D 6512 the lower scope limit was determined to be 9 g/L for chlorinated gold leaching barren effluent water.
This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific hazard statements are given in Section and Section .
2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.
ASTM Standards
D1129 Terminology Relating to Water
D1193 Specification for Reagent Water
D1293 Test Methods for pH of Water
D2036 Test Methods for Cyanides in Water
D2777 Practice for Determination of Precision and Bias of Applicable Test Methods of Committee D19 on Water
D3856 Guide for Management Systems in Laboratories Engaged in Analysis of Water
D4841 Practice for Estimation of Holding Time for Water Samples Containing Organic and Inorganic Constituents
D5847 Practice for Writing Quality Control Specifications for Standard Test Methods for Water Analysis
D6512 Practice for Interlaboratory Quantitation Estimate
D6696 Guide for Understanding Cyanide Species
D6888 Test Method for Available Cyanide with Ligand Displacement and Flow Injection Analysis (FIA) Utilizing Gas Diffusion Separation and Amperometric Detection
D7365 Practice for Sampling, Preservation and Mitigating Interferences in Water Samples for Analysis of Cyanide
E691 Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method
E1763 Guide for Interpretation and Use of Results from Interlaboratory Testing of Chemical Analysis Methods
Keywords
amperometry; aquatic free cyanide; diffusible cyanide, flow injection analysis; free cyanide; gas diffusion separation; Amperometric detection; Cyanides; Flow injection analysis (FIA); Gas diffusion ;
ICS Code
ICS Number Code 13.060.50 (Examination of water for chemical substances)
DOI: 10.1520/D7237-06