ASTM F1820-97(2009)
Historical Standard: ASTM F1820-97(2009) Standard Test Method for Determining the Axial Disassembly Force of a Modular Acetabular Device
SUPERSEDED (see Active link, below)
ASTM F1820
1. Scope
1.1 This test method covers a standard methodology by which to measure the attachment strength between the modular acetabular shell and liner. Although the methodology described does not replicate physiological loading conditions, it has been described as a means of comparing integrity of various locking mechanisms.
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.
ASTM Standards
E4 Practices for Force Verification of Testing Machines
Keywords
acetabular component; arthroplasty; disassembly; lever out; offset pullout; torque disassembly;
ICS Code
ICS Number Code 11.040.40 (Implants for surgery, prothetics and orthotics)
DOI: 10.1520/F1820-97R09
ASTM International is a member of CrossRef.